Boston Scientific NephroMax Kit, Material Numbers (UPN): 1. M0062101180; 2. M0062101600. (Urology) — Reason: The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter. — Status: Ongoing
Brand Name: MediHoney ¿ Product Name: MediHoney¿ Paste Model/Catalog Number: 31505 Product Description: MediHoney¿ Paste with Active Leptospermum Honey is a moist dressing made of 100% Active Leptospermum medical grade Honey. The dressing helps maintain a moist environment conducive to wound healing and helps aid and support autolytic debridement. The dressing can be removed without causing trauma to the wound when a primary dressing is used. — Reason: Packaging failures were identified which could lead to a breach in the sterile barrier. — Status: Ongoing
Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065750918 PAK,I/A LINE,NGVS 8065750957 ASSY,SHIP,CONSTELLATION VFC 8065750972 ENDOILLUMATOR,RFID,23G 8065751014 CONST AUTO GAS FILL PAK 8065751014 CONST AUTO GAS FILL PAK 8065751441 ENDOILLUMINATOR,RFID,25G 8065751577 CHANDELIER,25GA,W/RFID Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A — Reason: Due to incomplete seals in the pouch which provide the sterile barrier. — Status: Ongoing
t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor — Reason: A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia. — Status: Ongoing
Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Pinnacle TumorLOC 16.2 Model 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Pinnacle TumorLOC 16.2.1 Model 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Pinnacle TumorLOC 16.0.2 Model 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Pinnacle TumorLOC 14.0 Model 870200 UDI code: N/A Pinnacle3 Radiation Therapy Planning System is a software package intended to provide planning support for the treatment of disease processes. The full Pinnacle3 Radiation Therapy Planning System software package provides planning support for the treatment of disease processes, utilizing photon, electron and brachytherapy techniques. — Reason: Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations. — Status: Ongoing
Brand Name: CVS Honey Pad Product Name: CVS Honey Pad Model/Catalog Number: 9370613, 937404 Product Description: Infused with natural medical-grade honey Manuka (Leptospermum) honey is a medical-grade honey that is harvested from bees in New Zealand. Researchers have found that this species has unique components that make it ideal for wounds. — Reason: Packaging failures were identified which could lead to a breach in the sterile barrier. — Status: Ongoing
Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000252 UNITY 27 GA CHANDELIER 8065000253 UNITY 27+DS WIDE ANGLE LIGHT PROBE Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A — Reason: Due to incomplete seals in the pouch which provide the sterile barrier. — Status: Ongoing
Brand Name: MediHoney¿ Product Name: MediHoney¿ Adhesive Hydrogel Sheet Dressing Model/Catalog Number: 31720, 31740 Product Description: MEDIHONEY¿ Adhesive Hydrogel Sheet Dressing with Leptospermum Honey is a moist wound dressing made of Leptospermum Honey (+15) in combination with a hydrogel gelling agent, surrounded by an adhesive border. The gelling agent is contained within the dressing mass. The dressing contains 100% Leptospermum Honey. As wound exudate is absorbed, the dressing which has a low pH and high osmolarity forms a gel. This assists in maintaining a moist environment for optimal wound healing, aids in autolytic debridement, and allows for non-traumatic removal. This product is an occlusive dressing, and does not require a secondary dressing. — Reason: Packaging failures were identified which could lead to a breach in the sterile barrier. — Status: Ongoing
Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000393 UNITY HP VISCOUS FLUID CONTROL VFC PACK Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A — Reason: Due to incomplete seals in the pouch which provide the sterile barrier. — Status: Ongoing
Boston Scientific UroMax Ultra Kit, Material Numbers (UPN): 1. M0062251200, 2. M0062251210, 3. M0062251220, 4. M0062251230, 5. M0062251240, 6. M0062251260, 7. M0062251290, 8. M0062251300, 9. M0062251310. 10. M0062251350, 11. M0062251360, 12. M0062251370. (Urology) — Reason: The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter. — Status: Ongoing
Boston Scientific Encore 26 Inflation Device, Material Numbers (UPN): 1. H74904526011; 2. H74904526052; 3. M0067101140; 4. M001151050; 5. M001151062; 6. M00566670. (Interventional Cardiology, Peripheral Interventions) — Reason: The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter. — Status: Ongoing
Brand Name: MediHoney¿ Product Name: MediHoney¿ HCS Sheet Dressing with Leptospermum Model/Catalog Number: 31612, 31622, 31644 Product Description: MEDIHONEY¿ HCS SHEET DRESSING with LEPTOSPERMUM HONEY is a moist wound dressing made of Leptospermum Honey (+15) in combination with a hydrogel collodial gelling agent. The gelling agent is contained within the dressing mass. The dressing contains 63%, by weight, Leptospermum Honey. As wound exudate is absorbed, the dressing which has a low pH and high osmolarity forms a gel. This assists in maintaining a moist environment for optimal wound healing, aids in autolytic debridement, and allows for non-traumatic removal. This product is an occlusive dressing, and does require a secondary dressing. — Reason: Packaging failures were identified which could lead to a breach in the sterile barrier. — Status: Ongoing
Brand Name: MediHoney¿ Product Name: MediHoney¿ Hydrogel Model/Catalog Number: 31620, 31640 Product Description: MEDIHONEY¿ Non-Adhesive Hydrogel Sheet Dressing with Leptospermum Honey is a moist wound dressing made of Leptospermum Honey (+15) in combination with a hydrogel gelling agent. The gelling agent is contained within the dressing mass. The dressing contains 100%, Leptospermum Honey. As wound exudate is absorbed, the dressing which has a low pH and high osmolarity forms a gel. This assists in maintaining a moist environment for optimal wound healing, aids in autolytic debridement, and allows for non-traumatic removal. This product is an occlusive dressing, and does require a secondary dressing. — Reason: Packaging failures were identified which could lead to a breach in the sterile barrier. — Status: Ongoing
Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000250 UNITY 27+DS HYPERVIT 30K VIT PROBE 8065753168 UNITY 25+ 7.5K ANTERIOR VIT PROBE 8065753179 UNITY 25+ HYPERVIT 30K VIT PROBE Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A — Reason: Due to incomplete seals in the pouch which provide the sterile barrier. — Status: Ongoing
Brand Name: MediHoney ¿ Product Name: MediHoney ¿ Tube Model/Catalog Number: 31515 and 31535 Product Description: MediHoney¿ Paste with Active Leptospermum Honey is a moist dressing made of 100% Active Leptospermum medical grade Honey. The dressing helps maintain a moist environment conducive to wound healing and helps aid and support autolytic debridement. The dressing can be removed without causing trauma to the wound when a primary dressing is used. — Reason: Packaging failures were identified which could lead to a breach in the sterile barrier. — Status: Ongoing
Brand Name: MediHoney¿ Product Name: MediHoney¿ HCS Surgical Adhesive Sheet Model/Catalog Number: 31738 Product Description: MEDIHONEY¿ HCS SURGICAL ADHESIVE SHEET DRESSINGS with LEPTOSPERMUM HONEY is a moist wound dressing made of Leptospermum Honey (+15) in combination with a hydrogel colloidal gelling agent, surrounded by an adhesive border. The gelling agent is contained within the dressing mass. As wound exudate is absorbed, the dressing which has a low pH and high osmolarity forms a gel. This assists in maintaining a moist environment for optimal wound healing, aids and supports autolytic debridement, and allows for non-traumatic removal. The product is an occlusive dressing, and does not require a secondary dressing. — Reason: Packaging failures were identified which could lead to a breach in the sterile barrier. — Status: Ongoing
Brand Name: MediHoney¿ Product Name: MediHoney¿ Hydrogel Adhesive Sheet Model/Catalog Number: 31722, 31744 Product Description: MEDIHONEY¿ HCS ADHESIVE SHEET DRESSING with LEPTOSPERMUM HONEY is a moist wound dressing made of Leptospermum Honey (+15) in combination with a hydrogel colloidal gelling agent, surrounded by an adhesive border. The gelling agent is contained within the dressing mass. The dressing contains 63%, by weight, Leptospermum Honey. As wound exudate is absorbed, the dressing which has a low pH and high osmolarity forms a gel. This assists in maintaining a moist environment for optimal wound healing, aids in autolytic debridement, and allows for non-traumatic removal. This product is an occlusive dressing, and does not require a secondary dressing. — Reason: Packaging failures were identified which could lead to a breach in the sterile barrier. — Status: Ongoing
Boston Scientific Encore Advantage Kit, Material Numbers (UPN): 1. H74904527011; 2. H74904527052. (Interventional Cardiology and Peripheral Interventions) — Reason: The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter. — Status: Ongoing
Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000356 27GA HP ENTRY SYSTEM, 4 MM 1 COUNT 8065000358 25GA HP ENTRY SYSTEM, 6 MM 1 COUNT 8065000359 27GA HP ENTRY SYSTEM, 6 MM 1 COUNT 8065000467 25GA INFUSION CANNULA(4&6MM ES)+TRAY+TMD 8065000468 27GA INFUSION CANNULA(4&6MM ES)+TRAY+TMD Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A — Reason: Due to incomplete seals in the pouch which provide the sterile barrier. — Status: Ongoing
DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12110 LV VENT 10FR, Model Number 12110 — Reason: The catheters may not retain their shape. — Status: Ongoing
Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Nonvented stopper included. 50mL, Product Code 21-7001-24 — Reason: Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. — Status: Ongoing
Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy. — Reason: Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws. — Status: Ongoing
Smiths Medical CADD Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included. 100mL, Product Code 21-7302-24 — Reason: Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. — Status: Ongoing
Smiths Medical NRFit CADD Yellow Medication Cassette Reservoir with NRFit connector with flow stop, yellow-striped tubing, clamp and female NRFit connector. Yellow stopper included 100mL", Product Code 21-7600-24 — Reason: Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. — Status: Ongoing
ViziShot 2FLEX. Single Use Aspiration Needle. Model Number: NA-U403SX-4019. The ViziShot 2 FLEX tissue biopsy device is made up of four main parts: a handle, a sheath, a needle, and a stylet. The sheath and needle are attached to the handle, and the stylet, which is removable, is located inside the needle tube. — Reason: Potential for undetected, deformed a-traumatic tips. — Status: Ongoing
Reef TA Inserter, Part Number: TA2-001003, part of the Reef Interbody System — Reason: Inserter, used as part of a system with an intervertebral body fusion device, may contribute to difficulty disengaging the implant or improper implant positioning, which may may lead to prolonged surgery, intraoperative revision, and may require the use of alternate instrumentation. Devices may also have incorrect UDI-DI marking. — Status: Ongoing
Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Non-vented stopper included. 100mL, Product Code 21-7002-24 — Reason: Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. — Status: Ongoing
DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12115 LV VENT 15FR, Model Number 12115 — Reason: The catheters may not retain their shape. — Status: Ongoing
Smiths Medical CADD" Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included. 50mL, Product Code 21-7301-24 — Reason: Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. — Status: Ongoing
Xpert SA Nasal Complete Catalog Number: GXSACOMP-10 UDI-DI code: 07332940000158 Xpert SA Nasal Complete kit; 10 test kit: A collection of reagents and other associated materials intended to be used for the direct qualitative and/or quantitative detection of nucleic acid from methicillin-resistant and methicillin-susceptible Staphylococcus aureus (MRSA/MSSA) bacteria present in a clinical specimen, using a nucleic acid technique — Reason: As a result of Post-Market half-time stability testing, false negative test result was observed for a positive sample type. — Status: Ongoing
Smiths Medical CADD" Yellow Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included 100mL, Product Code 21-7300-24 — Reason: Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter. — Status: Ongoing
DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113 LV VENT 13FR, Model Number 12113 — Reason: The catheters may not retain their shape. — Status: Ongoing
FloPatch FP120, REF: FP120-FOT01-005 — Reason: Flowmeter patch as a human readable expiry date on the outer shipping packaging that reads 2025-06-20 when it should read 2026-06-20. — Status: Ongoing
Baster Continu-Flo Solution Set, Non-Vented, 2 Clearlink Luer Activated Valves, Backcheck Valve. 10 drops/mL. 112" (2.8 m), Product Code REF 2C8519; intravascular administration set — Reason: Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-Flo solution sets listed below due to customer reports of tubing separation. — Status: Ongoing
RayStation with the following product descriptions: 1. RayStation 11B, Software Version: 12.0.0.932. 2. RayStation 11B Service Pack 1, Software Version: 12.1.0.1221. 3. RayStation 11B Service Pack 2, Software Version: 12.0.3.68. 4. RayStation 11B Service Pack 3, Software Version: 12.0.4.12. 5. RayStation 11B Service Pack Toshiba 1, Software Version: 12.3.0.119. Product Description: Radiation Therapy Treatment Planning System. — Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. — Status: Ongoing
RayStation with the following product descriptions: 1. RayStation 8B, Software Version: 8.1.0.47. 2. RayStation 8B Service Pack 1, Software Version: 8.1.1.8. 3. RayStation 8B Service Pack 2, Software Version: 8.1.2.5. 4. RayStation 9A, Software Version: 9.0.0.113. 5. RayStation 9A Service Pack 1, Software Version: 9.0.1.142. Product Description: Radiation Therapy Treatment Planning System. — Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. — Status: Ongoing
IntelliSpace Cardiovascular, Software 8.0.0.4. — Reason: Software issue that results in the display of outdated information. — Status: Ongoing
Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006 — Reason: Vital sign monitoring instrument may trigger an error message during blood pressures measurements, Specifically, when the initial inflation pressure is set in the high-pressure range. — Status: Ongoing
Sterile Medline Convenience Kits: 1) EXTREMITY PACK, Model Number: DYNJ45701B — Reason: Convenience kits labeled as sterile have not gone through the sterilization process. — Status: Ongoing
RayStation with the following product descriptions: 1. RayStation 11A, Software Version: 11.0.0.951. 2. RayStation 11A Service Pack 1, Software Version: 11.0.1.29. 3. RayStation 11A Service Pack 2, Software Version: 11.0.3.116. 4. RayStation 11A Service Pack 3, Software Version: 11.0.4.15. Product Description: Radiation Therapy Treatment Planning System. — Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. — Status: Ongoing
RayStation with the following product descriptions: 1. RayStation 9B, Software Version: 9.1.0.933. 2. RayStation 9B Service Pack 1, Software Version: 9.2.0.483. 3. RayStation 10A, Software Version: 10.0.0.1154. 4. RayStation 10A Service Pack 1, Software Version: 10.0.1.52. 5. RayStation 10A Service Pack 2, Software Version: 10.0.2.10. Product Description: Radiation Therapy Treatment Planning System. — Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. — Status: Ongoing
RayStation with the following product descriptions: 1. RayStation 6 Service Pack 1, Software Version: 6.1.1.2. 2. RayStation 6 Service Pack 2, Software Version: 6.2.0.7. 3. RayStation 6 Service Pack 3, Software Version: 6.3.0.6. Product Description: Radiation Therapy Treatment Planning System. — Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. — Status: Ongoing
Sterile Medline Convenience Kits: 1) MAJOR PACK L-F, Model Number: DYNJ0382730O; 2) LB BASIC CUSTOM PACK, Model Number: DYNJ61038B — Reason: Convenience kits labeled as sterile have not gone through the sterilization process. — Status: Ongoing
RayStation with the following product descriptions: 1. RayStation 10B, Software Version: 10.1.0.613. 2. RayStation 10B Service Pack 1, Software Version: 10.1.1.54. Product Description: Radiation Therapy Treatment Planning System. — Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. — Status: Ongoing
RayStation with the following product descriptions: 1. RayStation 12A, Software Version: 13.0.0.1547. 2. RayStation 12A Service Pack 1, Software Version: 13.1.0.144. 3. RayStation 12A Service Pack 2, Software Version: 13.1.1.89. Product Description: Radiation Therapy Treatment Planning System. — Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. — Status: Ongoing
Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation sodium citrate 3.2%, REF: 454322 — Reason: Blood collection tubes were manufactured with the label applied too high on the tube, which shifted the printed fill level mark upward, which could result in unnecessary sample rejection and repeat collections. — Status: Ongoing
RayStation with the following product descriptions: 1. RayStation 7, Software Version: 7.0.0.19. 2. RayStation 8A, Software Version: 8.0.0.61. 3. RayStation 8A Service Pack 1, Software Version: 8.0.1.10. Product Description: Radiation Therapy Treatment Planning System. — Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. — Status: Ongoing
RayStation with the following product descriptions: 1. RayStation 2023B, Software Version: 14.0.0.3338. 2. RayStation 2024A, Software Version: 15.0.0.430. 3. RayStation 2024A SP3, Software Version: 15.1.3.10. Product Description: Radiation Therapy Treatment Planning System. — Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. — Status: Ongoing
Brand Name: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use (Dako Omnis) Model/Catalog Number: GA60461-2 Software Version: Not applicable Product Description: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use (Dako Omnis) is intended for use in immunohistochemistry together with the Dako Omnis instrument. This antibody labels cells of the B-cell lineage and is a useful aid for the classification of neoplasms of B-cell derivation. Component: Not applicable — Reason: Their is the potential for weak staining which may result in false negative CD20 identification. — Status: Ongoing
Brand Name: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use (Dako Omnis) Model/Catalog Number: GA60461-2J Software Version: Not applicable Product Description: FLEX Monoclonal Mouse Anti-Human CD20cy, Clone L26, Ready-to-Use (Dako Omnis) is intended for use in immunohistochemistry together with the Dako Omnis instrument. This antibody labels cells of the B-cell lineage and is a useful aid for the classification of neoplasms of B-cell derivation. Component: Not applicable — Reason: Their is the potential for weak staining which may result in false negative CD20 identification. — Status: Ongoing