device recall
saveRAYSEARCH LABORATORIES AB device recall (Cls II)
August 8, 2025updated 2mo ago
community read0 votes
RayStation with the following product descriptions: 1. RayStation 9B, Software Version: 9.1.0.933. 2. RayStation 9B Service Pack 1, Software Version: 9.2.0.483. 3. RayStation 10A, Software Version: 10.0.0.1154. 4. RayStation 10A Service Pack 1, Software Version: 10.0.1.52. 5. RayStation 10A Service Pack 2, Software Version: 10.0.2.10. Product Description: Radiation Therapy Treatment Planning System. — Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique. — Status: Ongoing
view source ↗