device recall
saveTandem Diabetes Care, Inc. device recall (Cls II)
community read0 votes
t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor — Reason: A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia. — Status: Ongoing
view source ↗Data sources
last refreshed 5mo ago
Found something off?
same company · ±60 days · 6
- device 510kTandem Diabetes Care — SteadiSet infusion set (510k cleared)Aug 6, 2025
- device recallTandem Diabetes Care, Inc. device recall (Cls I)Jul 21, 2025
- device pmaAbbott Diabetes Care, Inc. — FreeStyle Libre Pro Flash Glucose Monitoring System (PMA supplement)Jul 11, 2025
- device pmaAbbott Diabetes Care, Inc. — FreeStyle Libre 14-Day Flash Glucose Monitoring System (PMA supplement)Jul 11, 2025
- device recallTandem Diabetes Care, Inc. device recall (Cls II)Jul 3, 2025
- device 510kTandem Diabetes Care, Inc. — t:slim X2 insulin pump with interoperable technology (510k cleared)Sep 24, 2025
comments(0)
5-min edit window · permanent after that