pharmadog
News
when
  • Latest
  • Archive
by source
  • All Sources
  • Sources Page
Jobs
department
  • Clinical
  • Regulatory
  • Medical Affairs
  • Commercial
  • R&D / Discovery
  • Biostatistics / Data
  • Manufacturing / CMC
  • Market Access
therapeutic area
  • Oncology
  • Immunology
  • Neuroscience
  • Cardiovascular
  • Metabolic
  • Rare Disease
  • Infectious Disease
location & type
  • Remote Only
  • US Only
  • California
  • Massachusetts
  • Internships
  • Phase 3 Roles
  • All Jobs →
Sign InSubscribe
pharmadog

fetch the data · sniff the signal

Discover
  • Jobs
  • News
Hubs
  • Topics
  • Patent cliff
  • Publications
Tools
  • Compare
  • Search
  • Bookmarks
Trust
  • About
  • Sources
  • Contact
Legal
  • Privacy
  • Terms
  • Pricing

© 2026 pharmadog.xyz

made by humans and a good dog

  • home
  • jobs
  • news
  • search
fda medwatch·Jul 2·3 min read
save

Breathing Circuit Set Correction: Hamilton Medical Updates Use Instructions for Breathing Circuit Set

The ventilator breathing circuits are intended to connect a ventilator to a patient tube or mask. Obstruction may result in inadequate ventilation.

Jul 2, 2026·read at fda medwatch ↗

The affected products and recommendations for what to do with the devices below have not changed.This recall involves updating instructions for using certain devices, and does not involve removing them from where they are used or sold. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it without following the updated instructions.Affected Product The FDA is aware that Hamilton Medical has issued a letter to affected customers recommending certain coaxial breathing circuit sets used with HAMILTON-C1, HAMILTON T-1, and HAMILTON-MR1 ventilators have updated use instructions.Affected devices:Product NameBrand NameProduct NumberUDI-DIT1 CIRCUIT 240 +VALVE B/10Hamilton Breathing circuit sets, coaxial26012707630002802956T1 CIRCUIT 180 +VALVE B/20Hamilton Breathing circuit sets, coaxial26012807630002802963MR1 CIRCUIT 300 +VALVE B/10Hamilton Breathing circuit sets, coaxial26016707630002802970MR1 CIRCUIT 480 +VALVE B/8Hamilton Breathing circuit sets, coaxial26016807630002802987Full List of Affected LotsWhat to DoUsers should review and follow the pre-operative test instructions and troubleshooting instructions prior to continued use.On May 18, 2026, Hamilton Medical sent all affected customers a letter recommending the following actions:Verify unimpeded expiratory gas flow prior to use by ventilating a test lung using the settings provided by Hamilton Medical after performing the pre-operative test.If no “exhalation obstruction” alarm is issued, the expiratory valve is safe to use.If the “exhalation obstruction” alarm is triggered, do not use the coaxial breathing circuit set.

The coaxial breathing circuit set must be discarded and replaced.If the “Exhalation Obstructed” alarm is triggered on the ventilator, users should perform the following steps:Disconnect the patient from the ventilator and ensure alternative means of ventilation (e.g. alternative ventilator, hand bagging, etc.).Once patient safety is ensured, remove the expiratory valve set and overcome the membrane adhesion by detaching the membrane from the expiratory valve body once.Place the membrane back onto the body. Then, ensure the membrane is aligned with the expiratory valve body with the metal facing up (do not press onto the metal) and install the expiratory valve set into the exhaust port on the ventilator.Alternatively, use a new expiratory valve set.Inform all potential users of affected coaxial breathing circuit sets about this issue and post the provided notice in your facility.

Make sure recommended actions are followed.Affected products may continue to be used with the risk mitigations provided above.Reason for CorrectionHamilton Medical stated that certain lots of coaxial breathing circuit sets preassembled with an expiratory valve set do not perform as intended. In some cases, the expiratory valve membrane adhered to the sealing ring on the body of the expiratory valve, impairing valve opening and expiratory gas flow. The issue is not detected during the pre-operative test of the ventilator unless a test lung is actively ventilated.

If the expiratory valve membrane is adhered to the sealing ring, the malfunction appears within the first breaths after ventilation is initiated.When the expiratory gas flow is impeded, the end-expiratory pressure is too high, or the end-expiratory flow is too low. This will trigger the ventilator’s “Exhalation Obstructed” alarm.An obstruction of expiratory flow results in inadequate ventilation and impaired gas exchange. Acute respiratory compromise due to impaired exhalation may result in inadequate ventilation and oxygen desaturation (hypoxemia).As of May 13, Hamilton Medical has reported four serious injuries and no deaths associated with this issue.Device UseHamilton Medical, Inc.

ventilator coaxial breathing circuits are intended to connect HAMILTON-C1/T1/MR1 ventilators to a patient tube or respiratory mask during ventilation.Contact InformationCustomers in the U.S. with adverse reactions, quality problems, or questions about this issue should contact Hamilton Medical at complaints@hamiltonmedical.com or call 1-800-426-6331.Additional FDA Resources FDA Enforcement ReportCDRH Medical Device Recall DatabaseUnique Device Identifier (UDI) The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.

How do I recognize a UDI on a label?AccessGUDID database - Identify Your Medical DeviceBenefits of a UDI SystemHow do I report a problem? Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.Timeline of Communication UpdatesDateActions07/02/2026The FDA updated this communication to inform the public that this issue has been classified as a Class I Recall.06/08/2026The FDA issued an Early Alert communication to notify the public of a potentially high-risk device issue.

source

Reporting by fda medwatch.

read at fda medwatch ↗
731 words · retrieved 14m ago
sharex / twitterlinkedin

comments(0)

5-min edit window · permanent after that
sign in to leave a comment · permanent archive after 5 minutes
no comments yet — first sniff?

companies & drugs in this story

no entities indexed yet

related stories

  • 5d agoEarly Alert: Percutaneous Catheter Issue from Boston Scientificfda medwatch
  • Jul 20Early Alert: Intravascular Administration Set Issue from Baxterfda medwatch
  • Jul 17Catheter Introducer Kit Recall: Abiomed Removes 14Fr Low Profile Introducer Kitsfda medwatch
  • Jul 17Thoracic Stent Graft Recall: Bolton Medical Removes Relay Pro Systemfda medwatch
  • Jul 16Convenience Kit Recall: B. Braun Medical, Inc. Removes Spinal Anesthesia Kits Containing Huons Bupivacaine Hydrochloride in Dextrose Injection, USPfda medwatch