device recall
saveZap Surgical Systems device recall (Cls II)
July 25, 2025updated 2mo ago
community read0 votes
ZAP-X Radiosurgery System, Model: 300150 — Reason: If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table. — Status: Ongoing
view source ↗
comments(0)
5-min edit window · permanent after that