device recall
saveGE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING device recall (Cls II)
September 12, 2025updated 3mo ago
community read0 votes
Elscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging — Reason: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. — Status: Ongoing
view source ↗