Cook Medical Incorporated device recall (Cls I)
Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W ZTA2-P-42-173 ZTA2-P-42-173-W ZTA2-P-42-225 ZTA2-P-42-225-W ZTA2-P-44-125 ZTA2-P-44-125-W ZTA2-P-44-179 ZTA2-P-44-179-W ZTA2-P-44-233 ZTA2-P-44-233-W ZTA2-P-46-125 ZTA2-P-46-125-W ZTA2-P-46-179 ZTA2-P-46-179-W ZTA2-P-46-233-W ZTA2-PT-40-36-167 ZTA2-PT-40-36-167-W ZTA2-PT-40-36-217 ZTA2-PT-40-36-217-W ZTA2-PT-42-38-173 ZTA2-PT-42-38-173-W ZTA2-PT-42-38-225 ZTA2-PT-42-38-225-W ZTA2-PT-44-40-179 ZTA2-PT-44-40-179-W ZTA2-PT-44-40-233-W ZTA2-PT-46-42-179 ZTA2-PT-46-42-179-W ZTA2-PT-46-42-233 ZTA2-PT-46-42-233-W ZTA2-P-46-233 ZTA2-PT-44-40-233 ZTA2-P-40-217 ZTA2-P-40-117 ZTA2-P-40-167-DEMO ZTA2-P-44-152-W ZTA2-P-40-142 ZTA2-P-44-152 ZTA2-P-40-142-W — Reason: Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization. — Status: Ongoing
view source ↗
comments(0)
5-min edit window · permanent after that