saveSiemens Medical Solutions USA, Inc device recall (Cls I)
October 1, 2025updated 3mo ago
community read0 votes
MAGNETOM Cima.X Upgrade. Model Number: 11689304. — Reason: There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room. — Status: Ongoing