saveMedline Industries, LP device recall (Cls I)
November 26, 2025updated 3mo ago
community read0 votes
MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed, REF MDR107003E; 2) Full Electric Basic Homecare Bed, 4-pack,REF MDR107003E-4. — Reason: Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire. — Status: Ongoing