device recall
saveRAYSEARCH LABORATORIES AB device recall (Cls II)
November 28, 2025updated 3mo ago
community read0 votes
RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy Treatment Planning System — Reason: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. — Status: Ongoing
view source ↗
comments(0)
5-min edit window · permanent after that