device recall
saveMedtronic Inc device recall (Cls II)
community read0 votes
Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090 — Reason: Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration. — Status: Ongoing
view source ↗Data sources
last refreshed 5mo ago
Found something off?
same company · ±60 days · 6
- device pmaMedtronic, Inc. — Aurora EV ICD DVEA3E4 (PMA supplement)Dec 3, 2025
- device pmaMedtronic, Inc. — Micra AV TPS MC1AVR1 PNJ Micra AV2 TPS MC2AVR1 PNJ Micra TPS MC1VR01 PNJ Micra VR2 TPS MC2VR01 (PMA supplement)Dec 3, 2025
- device pmaMedtronic, Inc. — CareLink SmartSync Device Manager Base 24970A (PMA supplement)Dec 3, 2025
- device pmaMedtronic, Inc. — Adapta, Versa, Sensia IPG ADDR01, ADDRS1, ADSR03, SES01 LWP, NVZ Adapta, Versa, Sensia IPG SEDRL1 NVZ Adapta, Versa, Sen (PMA supplement)Dec 3, 2025
- device pmaMedtronic, Inc. — Adapta, Versa, Sensia IPG ADDRL1, ADDRS1, VEDR01, ADDR03, ADSR03, ADDR01, ADSR06, ADSR01, ADVDD01, SEDR01, ADDR06, SESR0 (PMA supplement)Dec 3, 2025
- device pmaMedtronic, Inc. — Aurora EV ICD (DVEA3E4) (PMA supplement)Dec 3, 2025