Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit. — Reason: A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes. — Status: Ongoing