Medtronic MiniMed, Inc. device recall (Cls II)
MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896) — Reason: All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. — Status: Ongoing
view source ↗same company · ±60 days · 6
- device recallMedtronic MiniMed, Inc. device recall (Cls II)Feb 13, 2026
- device recallMedtronic MiniMed, Inc. device recall (Cls II)Feb 13, 2026
- device pmaMedtronic Sofamor Danek USA, Inc. — Infuse Bone Graft (PMA supplement)Feb 13, 2026
- device recallMedtronic MiniMed, Inc. device recall (Cls II)Feb 13, 2026
- device recallMedtronic MiniMed, Inc. device recall (Cls II)Feb 13, 2026
- device recallMedtronic MiniMed, Inc. device recall (Cls II)Feb 13, 2026
comments(0)
5-min edit window · permanent after that