device recall
saveOlympus Corporation of the Americas device recall (Cls II)
February 25, 2026updated 3mo ago
community read0 votes
Brand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Number: PK-CF0533 Product Description: An assembly of devices intended to generate and apply radio-frequency (RF) alternating current to soft tissues for cutting and coagulation during an endoscopic or open surgical procedure. — Reason: Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use. — Status: Ongoing
view source ↗
comments(0)
5-min edit window · permanent after that