Access Thyroglobulin assay, a simultaneous one-step immunoenzymatic assay, Catalog Number 33860 — Reason: Access Thyroglobulin reagent lot 439163 may generate erroneously high patient results. Falsely increased Access Thyroglobulin results may lead a physician to pursue unnecessary diagnostic imaging studies and/or inappropriate therapy adjustments in patients being monitored for residual or recurrent thyroid cancer. — Status: Ongoing
Multi Joystick R-net. Electrical wheelchair component. — Reason: Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral. — Status: Ongoing
Blood Agar,5% Sheep Blood 100/PK, Product Number R01202 — Reason: The products may contain surface and subsurface contamination of Listeria monocytogenes. — Status: Ongoing
Plum Duo Infusion System, List Number: 400020401 — Reason: ICU Medical identified two sequences of programming events and alarm interactions that may cause the user interface to become unresponsive. — Status: Ongoing
BritePro Solo Single-Use Fiber Optic Mini Handle and Blade Ref: 040-02-0110U (Mac 1), 040-02-0120U (Mac 2), 040-02-0130U (Mac 3), 040-02-0131U (Mac 3 Strong Curve) 040-02-0140U (Mac 4), 040-02-0400U (Miller 00), 040-02-0410U (Miller 1), 040-02-0420U (Miller 2), 040-02-0430U (Miller 3), — Reason: Laryngoscope handles may not illuminate as intended. — Status: Ongoing
BritePro Solo Mini with Pro Miller Blade Size 1.5, REF:040-02-0415U — Reason: Laryngoscope handles may not illuminate as intended. — Status: Ongoing
BritePro Solo Single-Use Fiber Optic Laryngoscope Handle Refs:040-310AU and 040-310U — Reason: Laryngoscope handles may not illuminate as intended. — Status: Ongoing
Britepro Solo Stubby Single-Use Fiber Optic Laryngoscope Handle REF:040-03-0000U — Reason: Laryngoscope handles may not illuminate as intended. — Status: Ongoing
Blood/MacConkey Biplate 100/PK, Product Number R02049 — Reason: The products may contain surface and subsurface contamination of Listeria monocytogenes. — Status: Ongoing
Strep Selective II Agar, Product Number R01859 — Reason: The products may contain surface and subsurface contamination of Listeria monocytogenes. — Status: Ongoing
BritePro Solo Single-Use Fiber Optic Handle and Blade Refs: 040-331U (Mac 1), 040-332U (Mac 2), 040-333U (Mac 3), 040-335U (Mac 3.5) 040-334U (Mac 4), 040-341U (Miller 1), 040-342U (Miller 2), 040-343U (Miller 3), 040-344U (Miller 4), 040-361U (Phillips 1), 040-362U (Phillips 2) — Reason: Laryngoscope handles may not illuminate as intended. — Status: Ongoing
ELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS STEM EXTENSION, STRAIGHT, SPLINED, SLOTTED. Canal filling stem implants. — Reason: Unidentified substance of unknown impact was identified on devices during a retrospective evaluation. — Status: Ongoing
BritePro Solo Mini Single-Use Fiber Optic Laryngoscope Handle REF:040-309U — Reason: Laryngoscope handles may not illuminate as intended. — Status: Ongoing
Blood/EMB, Levine 100/PK, Product Number R02041 — Reason: The products may contain surface and subsurface contamination of Listeria monocytogenes. — Status: Ongoing
ELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS SEGMENTAL STEM CANAL FILLING STRAIGHT SLOTTED SPLINED 2/3 PLASMA. Canal filling stem implants. — Reason: Unidentified substance of unknown impact was identified on devices during a retrospective evaluation. — Status: Ongoing
ELEOS SEGMENTAL LIMB SALVAGE SYSTEM, ELEOS BOWED CANAL FILLING STEMS. Canal filling stem implants. — Reason: Unidentified substance of unknown impact was identified on devices during a retrospective evaluation. — Status: Ongoing
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ANTERIOR HIP PACK, MODEL/ITEM NUMBER AMAH40N; 2)TOTAL KNEE DAN MYER, MODEL/ITEM NUMBER CCKM67F; 3)HIP PROTHESIS PACK, MODEL/ITEM NUMBER HSTH03N; 4)DTC ANTERIOR HIP PACK (PS 209459), MODEL/ITEM NUMBER MIAH65B; 5)DR WHIDD0N TOTAL HIP CPP, MODEL/ITEM NUMBER MPWH01B; 6)KNEE ARTHROSCOPY, MODEL/ITEM NUMBER SPKA57T; 7)ANTERIOR HIP PACK, MODEL/ITEM NUMBER WVAH40H. — Reason: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met. — Status: Ongoing
Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, mg/dl, REF: 1004219; Pump, t:slim X2, Clinical Use Only, REF: 1004484; Pump, t:slim X2, Control-IQ, REF: 1005611; Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement, REF: 1005613; Pump, t:slim X2, Control-IQ, mg/dl, REF: 1005615; Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement, REF: 1005617; Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement, REF: 1006379; Pump, t:slim X2, Control-IQ 7.4, REF: 1006402; Pump, t:slim X2, Control-IQ 7.4, Replacement, REF: 1006406; Pump, t:slim X2, Control-IQ 7.4, REF: 1006408; Pump, t:slim X2, Basal-IQ, , REF: 1006419; Pump, t:slim X2, Basal-IQ, mg/dL, REF: 1006420; Pump, t:slim X2, Basal-IQ, mmoUL, Refurbished Replacement, REF: 1006535; Pump, t:slim X2, Basal-IQ, mg/dL, Refurbished Replacement, REF: 1006537; Pump, t:slim X2, Control-IQ, mg/dl , Replacement, REF: 1008255; Pump, t:slim X2, Control-IQ, mmoUL, Replacement, REF: 1008256; Pump, t:slim X2, Basal-IQ, mmoUL, Replacement, REF: 1008257; Pump, t:slim X2, Basal-IQ, mg/dl ,Replacement, REF: 1008258; Pump, t:slim X2, Basal-IQ, REF: 1010004; Pump, t:slim X2, Basal-IQ, Replacement, REF: 1010005; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010006; Pump, t:slim X2, Basal-IQ, Medicare, REF: 1010007; Pump, t:slim X2, Control-IQ, REF: 1010009; Pump, t:slim X2, Control-IQ Replacement, REF: 1010010; Pump, t:slim X2, Control-IQ, Medicare, REF: 1010011; Pump, t:slim X2, Control-IQ, Refurbished Replacement, REF: 1010012; Pump, t:slim X2, Control-IQ, REF: 1012867; Pump, t:slim X2, Control-IQ, Medicare, REF: 1012868; Pump, t:slim X2, Control-IQ, Replacement, REF: 1012869; Pump, t:slim X2, Control-IQ 7.8, REF: 1013195; Pump, t:slim X2, Control-IQ 7.8, Medicare, REF: 1013196; Pump, t :slim X2, Control-IQ 7.8, Replacement, REF: 1013197; Pump, t:slim X2, Control-IQ 7.8, Refurbished Replacement, REF: 1013198 — Reason: Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization. — Status: Ongoing
Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions. — Reason: Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel. — Status: Ongoing
Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS — Reason: Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the current infusion rate, the duration the pump has been running at this flow rate, and the magnitude of the rate change. The longer the duration the pump has been running at the current infusion rate and the larger the magnitude of the rate change, the larger the underinfusion that would be experienced. Additionally, Baxter has identified an increase in customer reports of over and underinfusion potentially due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Consistent with the instructions for use, customers should ensure that: 1) The door is fully open before loading the set. 2) The tubing is taut and loaded without slack in the pumping channel. — Status: Ongoing
Medline ReNewal Reprocessed St. Jude Medical Response and Supreme Diagnostic Electrophysiology Catheter Models 401150RH, 401206RH, 401207RH, 401210RH, 401211RH, 401212RH, 401222RH, 401223RH, 401226RH, 401227RH, 401228RH, 401260RH, 401261RH, 401271RH, 401305RH, 401306RH, 401309RH, 401310RH, 401311RH, 401312RH, 401317RH, 401353RH, 401357RH, 401381RH, 401392RH, 401399RH, 401400RH, 401425RH, 401430RH, 401433RH, 401434RH, 401435RH, 401436RH, 401438RH, 401441RH, 401442RH, 401445RH, 401443RH, 401444RH, 401448RH, 401449RH, 401450RH, 401451RH, 401453RH, 401466RH, 401468RH, 401474RH, 401475RH, 401859RH, 401860RH, 401863RH, 401864RH, 401865RH, 401876RH, 401877RH, 401878RH, 401890RH, 401891RH, 401892RH, 401893RH, 401966RH, 401967RH, 401979RH, 401994RH, 401996RH, 402004RH, 402008RH, 402009RH, 402010RH, 402011RH, 402012RH — Reason: Specific lots of reprocessed electrophysiology catheters may contain small residual particulates. — Status: Ongoing
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLBN11C; 2) LABOR & DELIVERY PACK, Model/Item Number LLLD19I. — Reason: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met. — Status: Ongoing
Spectrum IQ Infusion Pump, Product Code 3570009 — Reason: Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy. — Status: Ongoing
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) SFCH CRANI PACK (PS 132432), MODEL/ITEM NUMBER UICR10W; 2) SFCH SHUNT PACK (PS 139798), MODEL/ITEM NUMBER UISH55V. — Reason: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met. — Status: Ongoing
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANCV78BC; 2) VASCULAR PACK - 232024, MODEL/ITEM NUMBER ANVA82AB; 3) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48O; 4) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48P; 5) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18F; 6) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18G; 7) CABG ACCESSORY PACK, MODEL/ITEM NUMBER HGCB98F; 8) CV SET UP PACK, MODEL/ITEM NUMBER HGCV05K; 9) OPEN HEART PACK B - COMPONENTS, MODEL/ITEM NUMBER HGOH99H; 10) STRADLE TABLE PACK - 301659, MODEL/ITEM NUMBER MHCV10AM; 11) MINOR VASCULAR PACK, MODEL/ITEM NUMBER MTVS73D; 12) AV GRAFT RCH 0246952, MODEL/ITEM NUMBER RCAV40C; 13) OPEN HEART PART 1 RCH 0247159, MODEL/ITEM NUMBER RCOH61G; 14) CARDIO 612968, MODEL/ITEM NUMBER TNCA08T; 15) CV B PACK, MODEL/ITEM NUMBER TNCB10AM; 16) A-V FISTULA - 206042, MODEL/ITEM NUMBER UDAV16AD; 17) CAROTID PACK - 206045, MODEL/ITEM NUMBER UDCP24AF; 18) FEM POP PACK - 206052, MODEL/ITEM NUMBER UDFP36AH; 19) CARDIOVASCULAR SUPPLY (PS 907285), MODEL/ITEM NUMBER UICD62AN; 20) SFCH PEDS CARDIAC SUPPLY PACK (PS131653), MODEL/ITEM NUMBER UICS88S; 21) SFCH ECMO PACK (PS 107116), MODEL/ITEM NUMBER UIPD03Q; 22) THORACIC SETUP (PS 035700), MODEL/ITEM NUMBER UITC53AL; 23) VASCULAR PACK, MODEL/ITEM NUMBER UIVA75J; 24) CABG ACCESSORY PACK 175619, MODEL/ITEM NUMBER UMCA14E; 25) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42E; 26) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42F; 27) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42G. — Reason: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met. — Status: Ongoing
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) A-C SUPPLEMENT PACK - 205946, Model/Item Number ANCV67X; 2) KIDNEY PERFUSION PACK - 208439, Model/Item Number ANKP15AD; 3) LAPCHOLAPPY PACK, Model/Item Number FYLA38G; 4) CYSTO PACK, Model/Item Number HSCY13M; 5) PICC DRESSING CHANGE KIT, Model/Item Number HSDR10K; 6) MIDLINE CHANGE KIT, Model/Item Number HSMD20H; 7) SHOULDER ARTHOSCOPY PACK, Model/Item Number PDSA12D; 8) PDA LF SJW, Model/Item Number SWPD12K; 9) AAA PACK - 206038, Model/Item Number UDAA63AH; 10) BASIC CAB PACK - 206043, Model/Item Number UDBC35AF; 11) BASIC HEART PACK - 206044, Model/Item Number UDBH37AX; 12) FREE FLAP DRAPE PACK (PS 907864), Model/Item Number UIFF72AD; 13) PROCUREMENT PACK, Model/Item Number UIPC89D; 14) THYROID PACK (PS 096552), Model/Item Number UITR45W; 15) TRAUMA PACK 175634, Model/Item Number UMTR17B; 16) TRAUMA PACK 175634, Model/Item Number UMTR17C. — Reason: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met. — Status: Ongoing
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J; 3) LAP CHOLE PACK, Model/Item Number JKLC55H; 4) LAPAROSCOPY PACK, Model/Item Number LLLA21K; 5) GENERAL LAPAROSCOPY MPH, Model/Item Number MPGL15G; — Reason: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met. — Status: Ongoing
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAPAROSCOPY PACK, MODEL/ITEM NUMBER EVLP61I; 2) PELVISCOPY, MODEL/ITEM NUMBER EVPV29G. — Reason: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met. — Status: Ongoing
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO, Model/Item Number ASCY24A; 2) CYSTO PACK, Model/Item Number BUCY78E; 3) CYSTO PACK, Model/Item Number LMCP22T; 4) TUR BASIN PACK, Model/Item Number LMTU31T; 5) CYSTO, Model/Item Number SACY80S; 6) UROLOGY PACK, Model/Item Number SPUR64F. — Reason: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met. — Status: Ongoing
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) ROBOT HYST PACK, MODEL/ITEM NUMBER GRRB10E; 2) VAG HYST PACK, MODEL/ITEM NUMBER SPVH62P. — Reason: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met. — Status: Ongoing
3mensio Workstation (Vascular Fenestrated) software — Reason: When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage — Status: Ongoing
Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034 — Reason: Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur. — Status: Ongoing
Prosthesis Head B, 28 mm, 7 mm neck length. Item Number: 198-828/04. — Reason: Inconsistent size terminology and color coding used on labeling — Status: Ongoing
Plastic Trial Head Brown, 7 mm neck length. Item Number: 175-928/15. — Reason: Inconsistent size terminology and color coding used on labeling — Status: Ongoing
Pentax Medical Video Processor; Model Number: EPK-i8020c; — Reason: During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane. — Status: Ongoing
TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number: 82889. — Reason: Potential for missing pins at tip of inserter. — Status: Ongoing
Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA system. — Reason: The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results. — Status: Ongoing
SureStep Foley Tray System Bardex I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303316A — Reason: Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction — Status: Ongoing
SureStep Foley Tray System Lubri-Sil I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303414A — Reason: Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction — Status: Ongoing
Brand Name: ReQuest Measles IgM Product Name: Enzyme Linked Immunoabsorbant Assay, Rubeola IgM Model/Catalog Number: 01-190M Software Version: N/A Product Description: Measles IgM tests, performed using ELISA (Enzyme-Linked Immunosorbent Assay) method, used to detect the presence of IgM antibodies against measles virus in a patient's serum or plasma Component: not a component — Reason: Measles IgM Test Kit lacks premarket approval or clearance. — Status: Ongoing
IPC Powerease System REF 2300000 UDI-DI: 00613994448705 00643169406834. The IPC POWEREASE System is indicated for drilling, tapping, and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. The IPC POWEREASE System is also used in the placement of screws, or cutting of screws, posts, and rods. The IPC is indicated for the incision/cutting, removal, drilling, and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Cranial, Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures. — Reason: Due to out of the box wobble of the driver. — Status: Ongoing
SureStep Foley Tray System BARD¿ Lubri-Sil Foley Catheter Tray, REF: A942216 — Reason: Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction — Status: Ongoing
Orthopedic Pack, Model: 88041-12 AC2424601B 09/02/2029 Universal Pack, Model: 77165-06 AC2424902B 09/05/2029 — Reason: Surgical drape packs may have open seals, which may compromise the sterility of the product. — Status: Ongoing
Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software — Reason: Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) , — Status: Ongoing
Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER BOSS DRILL, 6.5mm Model/Catalog Number: 804-06-312 Product Description: Material: S.S., Non Sterile — Reason: Their is a potential that the reamer may kick or bind up during or immediately prior to use. — Status: Ongoing
Brand Name: DJO SURGICAL Product Name: FA S Altivate Reverse Glenoid Tray Model/Catalog Number: None Software Version: NA Product Description: Enovis" shoulder devices are intended for treatment of patients who are candidates for shoulder arthroplasty per the Indication for use. While total shoulder replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients. The AltiVate Reverse¿ Shoulder Prosthesis is indicated as a reverse shoulder replacement for patients with a functional deltoid muscle and a grossly deficient rotator cuff joint suffering from pain and dysfunction due to: " Severe arthropathy with a grossly deficient rotator cuff; " Previously failed joint replacement with a grossly deficient rotator cuff; " Fracture of glenohumeral joint from trauma or pathologic conditions of the shoulder including humeral head fracture, displaced 3- or 4-part fractures of proximal humerus, or reconstruction after tumor resection; " Bone defect in proximal humerus; " Non-inflammatory degenerative disease including osteoarthritis and avascular necrosis of the natural humeral head and/or glenoid; " Inflammatory arthritis including rheumatoid arthritis; " Correction of functional deformity. The glenoid baseplate is intended for cementless application with addition of screws for fixation. This device may also be indicated in the salvage of previously failed surgical attempts for anatomic and hemi procedures. All RSP¿ Monoblock and AltiVate Reverse¿ humeral stems are intended for cemented or cementless use. Enovis" instrumentation consists of devices and their accessories used in surgical procedures. Implantation of Enovis" products should only be performed with Enovis" instrumentation or instrumentation distributed by Enovis". Enovis" instruments and instrument cases are generally composed of titanium, stainless steel, aluminum, and/or polymeric materials. The cases may be multi-layered with various inserts to hold surgical instrumentation in place during handling and storage. The inserts may consist of trays, holders, and silicone mats. The instrument cases will allow sterilization of the contents to occur in a steam autoclave utilizing the cleaning, sterilization, and drying cycle that has been validated and detailed below. Instrument cases do not provide a sterile barrier and must be used in conjunction with FDA cleared sterilization wrap to maintain sterility. Instruments are provided non-sterile and should be stored in their original packaging until cleaned and sterilized according to the recommended guidelines listed below. Component: NA — Reason: Their is a potential that the reamer may kick or bind up during or immediately prior to use. — Status: Ongoing
Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD Model/Catalog Number: 804-06-311 Product Description: Material: S.S/Nitronic, Non-Sterile — Reason: Their is a potential that the reamer may kick or bind up during or immediately prior to use. — Status: Ongoing
Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER SLEEVE Model/Catalog Number: 804-06-310 Product Description: Material: S.S/CoCrMo, Non-Sterile — Reason: Their is a potential that the reamer may kick or bind up during or immediately prior to use. — Status: Ongoing
Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Model/Catalog Number: IMU7F-40, IMU7F-35, IMU7F-30, IMU7F-25, and IMU7F-20 Software Version: Not applicable Product Description: The Insightra Ultra IABP Catheter Kit is a 7 French catheter available with either a 40, 35, 30, 25, or 20 cc balloon. The catheter has two lumens and is designed to provide counterpulsation cardiac assist therapy. The outer lumen is a channel for helium used to inflate and deflate the balloon. The inner lumen is used for a guidewire and blood pressure measurement. It is packaged as a convenience kit with accessories (one-way valve, 50 cc syringe, guide wires, introducer needle, dilator, pressure tubing, Datascope & Arrow adaptors). The balloon inflation and deflation cycle is synchronized (using a commercially available control console) with the ECG or arterial pressure to provide counterpulsation synchronized with the heartbeat. It is intended to increase coronary perfusion, decrease the workload of the left ventricle and allow healing of the myocardium Component: No — Reason: Device contains indications for use and device compatibility claims that have not be reviewed and approved for safety and effectiveness by the FDA. Instructions for use are not consistent with product training. — Status: Terminated
STA-Liquid Anti-Xa 4. Model/Catalog Number: 00311US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite analyzers, for the quantitative determination of the plasma levels of unfractionated (UFH) and low molecular weight (LMWH) heparins by measuring their anti-Xa activity on antithrombin in a competitive assay using a synthetic chromogenic substrate. — Reason: Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications. — Status: Ongoing