Halyard TRANSPORT BAG KIT. Kit Code: LIFE0080-01. — Reason: Tyvek bag seal issue which may compromise sterility of the kit. — Status: Ongoing
ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 267-01-000-10-N UDI-DI code: 00190376541471; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker. — Reason: Due a design issue where the navigated array connection geometry is incorrect. — Status: Ongoing
ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 266-01-000-10-N UDI-DI code: 00190376476179; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker. — Reason: Due a design issue where the navigated array connection geometry is incorrect. — Status: Ongoing
VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebactam (ipr01n); CARD NAME/ PRODUCT REFERENCE: AST-N399 423701 AST-N402 423644 AST-N403 423645 AST-N404 423664 AST-N405 423864 AST-N406 423869 AST-N407 423870 AST-N409 423925 AST-N412 423936 AST-N413 423928 AST-N414 423933 AST-N415 423934 AST-N416 423935 AST-N417 423880 AST-N419 423948 AST-N421 424055 AST-N428 424320 AST-N429 424321 AST-N430 424322 AST-N433 424389 AST-N439 424501 AST-N440 424502 AST-N443 424541 AST-N444 424587 AST-N467 424857 AST-N476 424934 AST-N478 425052 AST-N480 425084 AST-N481 425085 AST-N485 425182 AST-N496 425265 AST-N800 423310 AST-N802 423706 AST-N804 424634 AST-N808 424711 AST-N810 424712 AST-N811 424713 AST-N813 424722 AST-N815 425054 AST-XN09 423425 AST-XN15 423829 AST-XN17 423673 AST-XN18 423874 AST-XN20 423947 AST-XN21 424197 AST-XN22 424199 AST-XN23 424198 AST-XN24 424351 AST-XN26 424500 AST-XN28 424586 AST-XN29 424604 AST-XN30 424639 AST-XN31 424640 AST-XN32 424678 AST-XN33 424723 AST-XN34 424752 AST-XN35 424810 AST-XN36 424860 AST-XN37 424888 AST-XN38 425074 AST-XN39 425086 AST-XN40 425095 AST-XN41 425204 AST-XN42 425309 — Reason: Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa with the identified antibiotic formulations. Issue may lead to QC failure or false resistant results. — Status: Ongoing
Raz-AT (Attendant Tilt) Mobile Shower Commode Chair, Catalog Number Z300 UDI-DI code: (01)00628758003007 Raz-ART (Attendant Rotational Tilt) Mobile Shower Commode Chair, Catalog Number Z333 UDI-DI code: (01)00628758003335 Raz-AT600 (Bariatric Tilt) Mobile Shower Commode Chair, Catalog Number Z360 UDI-DI code: (01)00628758003601 Raz-SP (Self Propel) Mobile Shower Commode Chair, Catalog Number Z200 UDI-DI code: (01)00628758001003 Raz-AP (Attendant Propel) Mobile Shower Commode Chair, Catalog Number Z100 UDI-DI code: (01)00628758001003 The Raz Mobile Shower Commode Chair is intended to be used by people with mobility impairments, who may have difficulty in standing or keeping their balance while completing activities such as showering, toileting, or personal hygiene. — Reason: Due to a modification by a vendor and may be reinstalled without properly engaging the outside rear seat brackets that are used to keep the seat in place. — Status: Ongoing
Medline Convenience Kits: 1) NAIL KIT, Model Number: DYKM1528 — Reason: Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products. — Status: Ongoing
Meijer STERILE saline wound wash, First Aid Cleansing Spray, 0.9% USP SODIUM CHLORIDE, 7.1FL OZ (210 mL), DBD-CAN,SPRAY,7.1-OZ,SALINE,WOUND WASH,S, Model Number MJSALINE7 — Reason: Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products. — Status: Ongoing
EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705 — Reason: 942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery. — Status: Ongoing
Medline Sterile Saline Wound Wash, Cleansing & Wound Irrigation, USP Normal Saline (0.9% Sodium Chloride), 7.1 FL OZ (210 mL), DBD-CAN,SPRAY,7.1-OZ,SALINE,WOUND WASH,S, Model Number MDSALINE7 — Reason: Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products. — Status: Ongoing
Medline Convenience Kits: 1) TRUNK KIT W/EXPIRATION, Model Number: DYKM1361A; 2) TRUNK KIT 1EA, Model Number: DYKM2013; 3) TRUNK KIT 1EA, Model Number: DYKM2013A; 4) RN TRUNK KIT, Model Number: DYKM2699; 5) TRUNK KIT, Model Number: DYKTRUNK1 — Reason: Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products. — Status: Ongoing
EMPOWR Acetabular System: Liner, 10 degree Hooded, HXe+, 36H, REF: 942-01-36H; Liner, 10 degree Hooded, HXe+, 36J, REF: 942-01-36J; Liner, 10 degree Hooded, HXe+, 40G, REF: 942-01-40G; Liner, 10 degree Hooded, HXe+, 32C, REF: 942-01-32C — Reason: 942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery. — Status: Ongoing
Medline Convenience Kits: 1) NAIL KIT, Model Number: POD14214 — Reason: Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products. — Status: Ongoing
PRIMSAFLEX M100 SET, Product Code 106697; Dialyzer, High Permeability With Or Without Sealed Dialysate System — Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit. — Status: Ongoing
PRISMAFLEX ST100 SET US, Product Code 107636US; Dialyzer, High Permeability With Or Without Sealed Dialysate System — Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit. — Status: Ongoing
OXIRIS SET, Product Code 112016; Dialyzer, High Permeability With Or Without Sealed Dialysate System — Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit. — Status: Ongoing
PRISMAFLEX M150 SET, Product Code 109990; Dialyzer, High Permeability With Or Without Sealed Dialysate System — Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit. — Status: Ongoing
PRISMAFLEX TPE2000 SET, Product Code 114093; Dialyzer, High Permeability With Or Without Sealed Dialysate System — Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit. — Status: Ongoing
PRISMAFLEX HF1000 SET, Product Code 107140; Dialyzer, High Permeability With Or Without Sealed Dialysate System — Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit. — Status: Ongoing
PRISMAFLEX ST150 SET US, Product Code 107640US; Dialyzer, High Permeability With Or Without Sealed Dialysate System — Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit. — Status: Ongoing
WB Impression Post Closed Tray incl. guide screw, L 13mm, TAN/POM — Reason: The mix up of the impression caps provided in the package. The provided impression caps are magenta instead of brown. — Status: Ongoing
PRISMAFLEX HF1000 SET, Product Code 107142; Dialyzer, High Permeability With Or Without Sealed Dialysate System — Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit. — Status: Ongoing
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number: DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number: DYNDH1143C; 4) BIOPSY SPECIALS PACK, Model Number: DYNDH1873; 5) BIOPSY SPECIALS PACK, Model Number: DYNDH1934; 6) BREAST BIOPSY PACK, Model Number: DYNJ17489I; 7) VOR PACK-LF, Model Number: DYNJ48539; 8) VOR PACK-LF, Model Number: DYNJ48539A; 9) IR BIOPSY PACK, Model Number: DYNJ59897B; 10) BREAST PACK, Model Number: DYNJ82998 — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Medline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0330; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Brilliance iCT; Product Code (REF): 728306; — Reason: As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled. — Status: Ongoing
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0725; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Medline Convenience Kits: 1) KIT ACES INSERT PEG, Model Number: DYKMBNDL130; 2) KIT ACES INSERT PEG, Model Number: DYKMBNDL130A; 3) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4; 4) NICU LINE CHANGE TRAY, Model Number: LC295; 5) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977C; 6) PICC CATHETER INSERTION TRAY, Model Number: CVI4300 — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Medline Convenience Kits: 1) DRAPE PACK-CHOICE, Model Number: DYNJ63118A — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Spectral CT; Product Code (REF): 728333; — Reason: As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled. — Status: Ongoing
Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43 — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Medline Convenience Kits: 1) ADMIT KIT, Model Number: DYKA1343A — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3025 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Medline Convenience Kits: 1) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407N; 2) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431K; 3) TOT/ GYN LITHOTOMY CDS, Model Number: CDS985431O; 4) KIT ROBOTICS GYN, Model Number: DYKMBNDL109A; 5) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116A; 6) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116G; 7) ROBOTIC PACK, Model Number: DYNJ04135B; 8) MINOR LITHOTOMY PACK, Model Number: DYNJ04382J; 9) NBMC D & C PACK-LF, Model Number: DYNJ0665843C; 10) D&C PACK-LF, Model Number: DYNJ22567G; 11) D&C PACK-LF, Model Number: DYNJ22567J; 12) D AND C, Model Number: DYNJ41272B; 13) D AND C, Model Number: DYNJ41272C; 14) OB PACK, Model Number: DYNJ42884; 15) D&C/GYN PACK, Model Number: DYNJ47713A; 16) PERI GYN PACK, Model Number: DYNJ55377D; 17) ANTERIOR CERVICAL-LF, Model Number: DYNJ58344; 18) D&C PACK, Model Number: DYNJ67214D; 19) D&C PACK, Model Number: DYNJ67214F; 20) D&C PACK, Model Number: DYNJ67214G; 21) D&C PACK, Model Number: DYNJ67214I; 22) LITHOTOMY PACK, Model Number: DYNJ80765C; 23) GYN PACK, Model Number: DYNJ83475; 24) D AND C PACK, Model Number: DYNJ87468; 25) D&C-ASC, Model Number: DYNJ900465C; 26) D&C-ASC, Model Number: DYNJ900465D; 27) D&C-ASC, Model Number: DYNJ900465F; 28) GYN LITHOTOMY-LF, Model Number: DYNJ902711L; 29) GYN LITHOTOMY-LF, Model Number: DYNJ902711N; 30) D&C, Model Number: DYNJ906947C; 31) PACK PERI GYN DSMC, Model Number: DYNJT6424 — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-431Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0725; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0730 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, Model Number: CDS860015W; 4) MAJOR LAPAROTOMY CDS, Model Number: CDS860015X; 5) MINOR LAPAROTOMY CDS, Model Number: CDS860016P; 6) MINOR LAPAROTOMY CDS, Model Number: CDS860016Q; 7) MINOR LAPAROTOMY CDS, Model Number: CDS860016R; 8) MINOR LAPAROTOMY CDS, Model Number: CDS860016S; 9) LAP GASTRIC BYPASS CDS, Model Number: CDS860146M; 10) NEW LONDON MINOR CDS, Model Number: CDS982518M; 11) NEW LONDON MINOR CDS, Model Number: CDS982518N; 12) NEW LONDON MINOR CDS, Model Number: CDS982518O; 13) NEW LONDON MINOR CDS, Model Number: CDS982518P; 14) GENERAL LAPAROSCOPY CDS, Model Number: CDS983311J; 15) LAVH CDS, Model Number: CDS983411G; 16) NHP ROBOTIC GENERAL CDS, Model Number: CDS984244J; 17) ROBOTIC-LF, Model Number: CDS984543C; 18) CYSTO CDS, Model Number: CDS984592G; 19) CYSTO CDS, Model Number: CDS984592I; 20) ROBOTIC, Model Number: CDS985270F; 21) GENERAL LAPAROSCOPY CDS, Model Number: CDS985422I; 22) ROBOTIC CDS, Model Number: CDS985429L; 23) LAP CHOLE CDS, Model Number: CDS985558N; 24) LAP CHOLE CDS, Model Number: CDS985558O; 25) ENDO KIT, Model Number: DYKE1580A; 26) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105; 27) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105A; 28) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105B; 29) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116; 30) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116B; 31) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117A; 32) KIT SURG ONC DIAGNOSTIC LAPARO, Model Number: DYKMBNDL12; 33) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121; 34) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121A; 35) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121B; 36) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121C; 37) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121F; 38) KIT L&D CYSTO SETUP, Model Number: DYKMBNDL135A; 39) KIT THORACIC LAPAROSCOPIC HIAT, Model Number: DYKMBNDL136; 40) KIT THORACIC LAPAROSCOPIC HIAT, Model Number: DYKMBNDL136A; 41) KIT THORACIC STAGING LAPAROTOM, Model Number: DYKMBNDL15; 42) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153A; 43) KIT SURG ONC LAP DIAG/PORT INS, Model Number: DYKMBNDL153B; 44) KIT LAP DONOR NEPHRECTOMY, Model Number: DYKMBNDL173; 45) KIT LAP DONOR NEPHRECTOMY, Model Number: DYKMBNDL173A; 46) KIT THORACIC HIATAL HERNIA TRA, Model Number: DYKMBNDL175; 47) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176; 48) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176A; 49) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176B; 50) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176C; 51) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184; 52) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184A; 53) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188A; 54) KIT UROLOGY LAP SIDDIQUI, Model Number: DYKMBNDL193; 55) KIT UROLOGY LAP SIDDIQUI, Model Number: DYKMBNDL193A; 56) KIT UROLOGY BUCCAL GRAFT, Model Number: DYKMBNDL206; 57) KIT STC PLASTIC FREE FLAP, Model Number: DYKMBNDL54A; 58) KIT UROLOGY CYSTOSCOPY SUPPLEM, Model Number: DYKMBNDL60; 59) KIT UROLOGY CYSTOSCOPY SUPPLEM, Model Number: DYKMBNDL60A; 60) KIT UROLOGY MINOR, Model Number: DYKMBNDL61A; 61) KIT UROLOGY MINOR, Model Number: DYKMBNDL61B; 62) KIT GEN SURG LAP COLECTOMY, Model Number: DYKMBNDL73; 63) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73A; 64) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73B; 65) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73C; 66) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73D; 67) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74; 68) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74A; 69) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74B; 70) KIT GEN SURG LAP GASTRIC BYPAS, — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0720; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Medline Convenience Kits: 1) KIT GENERAL CLOSURE, Model Number: DYKMBNDL200A — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Medline Convenience Kits: 1) KIT STC TRACHEOSTOMY, Model Number: DYKMBNDL55; 2) KIT STC TRACHEOSTOMY, Model Number: DYKMBNDL55A; 3) TRACH TOTE, Model Number: DYNJ85691 — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model Number: CDS983782C; 4) BAPTIST FLOYD BRONCHOSCOPY, Model Number: DYKE1456D; 5) BRONCHOSCOPY SETUP KIT, Model Number: DYKE1955; 6) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156; 7) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156A; 8) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156B; 9) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174; 10) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174A; 11) KIT ENT LARYNGECTOMY, Model Number: DYKMBNDL174B; 12) KIT PEDS HEAD AND NECK, Model Number: DYKMBNDL196; 13) KIT OMF LEFORTE, Model Number: DYKMBNDL1A; 14) KIT OMF LEFORTE, Model Number: DYKMBNDL1B; 15) KIT OMF LEFORTE, Model Number: DYKMBNDL1C; 16) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32; 17) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32A; 18) KIT ENT ENDOSINUS GENERIC, Model Number: DYKMBNDL32B; 19) KIT GEN SURG THYROIDECTOMY, Model Number: DYKMBNDL66A; 20) KIT GEN SURG THYROIDECTOMY, Model Number: DYKMBNDL66B; 21) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88A; 22) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88B; 23) KIT GEN SURG EXCISION THYROID, Model Number: DYKMBNDL88C; 24) KIT GEN SURG LAP VENTRA, Model Number: DYKMBNDL96C; 25) MINOR ENT PACK, Model Number: DYNJ01831I; 26) ENT BASIC PACK-LF, Model Number: DYNJ0387711F; 27) D AND C PACK-LF, Model Number: DYNJ0855090P; 28) CSS SINUS PACK, Model Number: DYNJ17005I; 29) CSS HEAD & NECK PACK, Model Number: DYNJ26780I; 30) NASALPLASTY #61-RF, Model Number: DYNJ27320V; 31) MASTOID PACK-WINTER PARK-LF, Model Number: DYNJ32060C; 32) T&A PACK, Model Number: DYNJ32809I; 33) HEAD AND NECK PACK, Model Number: DYNJ41169; 34) SIMPLE NECK DISSECTION, Model Number: DYNJ47481C; 35) ENT PACK, Model Number: DYNJ47882F; 36) ENT PACK, Model Number: DYNJ48401C; 37) ENT PACK, Model Number: DYNJ51934A; 38) PK,ORTHO-EXTREMITY-LOSROBLES, Model Number: DYNJ52916A; 39) HEAD & NECK PACK, Model Number: DYNJ54964A; 40) HEAD & NECK PACK, Model Number: DYNJ54964B; 41) ENT - MINOR ENT PACK-LF, Model Number: DYNJ56840B; 42) GLENNON HEAD AND NECK PACK-LF, Model Number: DYNJ57200C; 43) ENT PACK, Model Number: DYNJ57688F; 44) RF T AND A PACK, Model Number: DYNJ61393A; 45) T & A PACK, Model Number: DYNJ61935D; 46) ENT PACK, Model Number: DYNJ62093B; 47) ENT PACK, Model Number: DYNJ64024B; 48) OSCLJ HEAD AND NECK PACK, Model Number: DYNJ66509A; 49) HEAD AND NECK PACK, Model Number: DYNJ67208; 50) HEAD & NECK PACK, Model Number: DYNJ67328B; 51) PEDS HEAD AND NECK PACK, Model Number: DYNJ69746B; 52) NASAL PACK, Model Number: DYNJ81280A; 53) ENT-PLASTICS PACK, Model Number: DYNJ84258A; 54) ENT-PLASTICS PACK, Model Number: DYNJ84258B; 55) HEAD & NECK PACK, Model Number: DYNJ85078A; 56) ENT PACK, Model Number: DYNJ86194A; 57) EENT PACK, Model Number: DYNJ86235; 58) ENT PACK, Model Number: DYNJ87378; 59) ENT PACK, Model Number: DYNJ87378A; 60) ENT PACK, Model Number: DYNJ87378D; 61) ENT SPLIT PACK, Model Number: DYNJ87451; 62) ENT SINUS PACK, Model Number: DYNJ88498; 63) NASAL, Model Number: DYNJ902993F; 64) ENDO SINUS-MIDTOWN, Model Number: DYNJ905866; 65) T A, Model Number: DYNJ906292B; 66) THYROID, Model Number: DYNJ906841A; 67) THYROID - SMH AMB, Model Number: DYNJ907739B; 68) ENT, Model Number: DYNJ911623; 69) HEAD AND NECK PACK, Model Number: DYNJT5739; 70) RO MEDIASTINAL PACK-LF, Model Number: PHS393087007C — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A; — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0730; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing
Medline Convenience Kits: 1) PPE KIT, Model Number: DYK1011945P — Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. — Status: Ongoing
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0720 Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current. — Reason: Devices which did not undergo thermoforming could deform and lose performance. — Status: Ongoing