Supplement type: 30-Day Notice. Product code: OZP. Advisory committee: CH. Decision: OK30
BVI Leos Laser and Endoscopy System. Model Number: OME6000U. — Reason: Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable). — Status: Ongoing
Revogene C. difficile, in vitro diagnostic assay intended for use with the Revogene instrument, Catalog Number 410300 — Reason: The affected lots show a decline in performance over time, which may lead to false-negative results. — Status: Ongoing
LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471 — Reason: A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ programs (OGPs) the collimation may be incorrect. — Status: Ongoing
Brand Name: LimFlow Vector Product Name: LimFlow Vector Model/Catalog Number: VT-US-23 — Reason: Incorrect expiration date — Status: Ongoing